FONA XPan 3D Operating Instructions
16/28
69 683 70110
-
150831
7.
Operation
7.1 Preparing for exposure
A.
Switching ON the Unit
ATTENTION
Following extreme temperature fluctuations, condensate formation
may occur; therefore please do not switch on the device until normal
room temperature has been reached.
By pressing the mains switch in the lower part of the vertical carriage
under the mirror, the unit is supplied as indicated by the green light of
the mains switch.
The display on the control panel turns on too
System initialization is started
Start of the reset function has to be performed
ATTENTION
When switching on the unit there must NOT be a patient positioned in
the unit.
If a fault occurs which requires switching the unit off and then back
on again, the patient must be taken out of the unit at the latest
before switching it on again!
B.
Reset Function
By pressing the RETURN Arm Movement key the rotation arm locates
the reference points and moves to the PATIENT ENTRY position, with
control panel and X-ray source to the right of the patient, image
receiver to the left.
C.
Switching ON the PC
Prepare the OrisWin DG Suite program on PC in stand-by for exposure
(see §5, Application Software).
* Do Pan Reset *
67 6.3
* Do Pan Reset *
67 6.3
Initialization
Please wait: xx
Device code:
110X1