21
Operating instructions - mobiSERVE
GB
Example:
Percent error occurred E %
-0,9 %
Current calibration factor
1.000
New USER K FACTOR
1.000 x {[100-(-0,9)]÷100}=
1.000 x [(100+0,9)÷100]=
1.009
If the counter indicates less than the real discharge value (negative error), the new calibration factor must
be higher than the previous one, as the example shows. Vice versa, if the counter indicates more than the
real discharge value (positive error).
Operation
Display Configuration
1
NONE
Counter in normal mode, not counting
2
LONG PRESS ON THE TOTAL KEY
The counter enters into the calibration mode and the
calibration factor used is displayed instead of the partial
amount. The messages FACT or USER on the screen indicate
which one of the two factors is currently being used (user
or manufacturer).
3
LONG PRESS ON THE RESET KEY
The counter indicates the TOTAL screen and the resettable
total amount indicator is zero.
The counter is ready to carry out the calibration by
discharge.
4
LONG PRESS ON THE RESET KEY
Direct change of the calibration factor. The message
DIRECT and the calibration factor currently being used ap-
pear on the display. On the lower left corner of the screen,
an arrow is shown, indicating the direction in which the
value displayed is changed. Increase or reduce when the
operations 5 or 6 are carried out.
5
SHORT PRESS ON THE RESET KEY
Change of arrow direction. The operation can be repeated
in order to change the arrow direction.