20
GB
Operating instructions - mobiSERVE
8
LONG PRESS ON THE RESET KEY
The mobiSERVE is informed that the calibration procedure
is finished.
Before this operation is carried out, check to ensure
that the STANDARD VALUE corresponds to the ACTUAL
VALUE.
Actual value
The mobiSERVE calculates the new USER K FACTOR. This
calculation may take some seconds, according to the
correction to be made. During this procedure, the arrow
disappears, but TOTAL remains on the screen.
If this procedure is carried out after point 5 without
changing the value indicated, the USER K FACTOR is equal
to the FACTORY K FACTOR and is thus ignored.
9
NO OPERATION
After the calculation, the new USER K FACTOR is indicated
for some seconds, then, the restart procedure is repeated in
order to finally get into the standby mode.
ATTENTION: From that moment, the value indicated is the
new calibration factor used by the counter, and it remains
so, also after a battery replacement!
10
NO OPERATION
The mobiSERVE stores the new operating calibration factor
and is prepared for discharge with the USER K FACTOR just
calculated.
9.4.3 Direct change of the K FACTOR
This procedure is especially helpful in order to correct an average error, which may occur because of the
multitude of tasks carried out. If, during normal counter operation, an average percent error occurs, this
one may be corrected by altering the calibration factor currently being used by the same percent value. In
this case, the percentage of the correction of the USER K FACTOR can be calculated by the user as follows:
new calibration factor = old calibration factor x