Operating manual BMGZ 600A
19
6.5
Wiring of the Connection Cable
Wiring between measuring roller and evaluation electronic unit must be done with
the shipped 8-wired, shielded twisted-pair cable (4x2x0.75mm
2
). Length of the
cable is done according to customer specification (specify when ordering). The
cable must be installed separate from power lines to prevent any inductive
disturbances.
On the measuring roller
side, the plastic sheath will
be removed on a length of
about 14cm. The white wire
is not used. The shield will
be connected to terminal 5.
On the electronic unit side,
the plastic sheath will be
removed on a length of
about 25-54cm. The white
wire is not used. The shield will be connected to the PG gland referring to fig. 14.
The contacts will be made referring to fig. 15 or wiring table. The shield of the
cable must be connected on both sides.
If a housing version „K“ (with steel cabinet) is used, the cable
will be led through the
steel cabinet using a 8-pole plug („13.3 List of parts“, Pos. 64).
fig. 15: Wiring of the measuring roller to the evaluation electronic unit B610002e
fig. 14: Shield connection to the electronic unit
B600024e