11
VIBRONET Installation 06.2016
Technical data
PARAMETER
VIB 5.802
Interfaces
Meas. channels, analog
6 differential inputs (3 of them synchronous)
Meas. channels, digital
RPM / Counter: 2 x TTL...30 V
Keyphaser: ± 30 V AC and DC
Input channels, digital
4 x, TTL...30 V
Output channels, digital
8 x, 5 V, 5 mA
Ethernet
1, data rate: 100 Mbit
Serial - RS 232
2, data rate: 115,2 kBit
FET switching output
12 V DC, 1 A, switchable
Expanded no. of channels
External multiplexer for analog and digital outputs
Measur
ement
Meas. range, analog
±10 V, ±1 V, ±100 mV, ±10 mV
Dynamic Range / Resolution
96 dB / 16 bit ADC
Accuracy, analog input
0.05% of full scale
Common mode rejection
> 115 dB at an amplification of 60 dB
Temperature coefficient, analog input
20 ppm / K
Input protection
Differential input: ± 12 V
Digital input: + 30 V
SW-Downsampling
4.8 / 2.4 / 1.2 / 0.6 / 0.3 / 0.15 kHz
Phase error, synchronous analog inputs
< 0,05 %
Crosstalk between analog inputs
< -100 dB
Dynamic amplitude errors
< -0.1 dB (up to 50% of the max. signal frequency)
< - 0.5 dB (up to 75% of the max. signal frequency)
< -1.0 dB (up to 80% of the max. signal frequency)
< - 3.0 dB (up to 100% of the max. signal frequency)
Counter frequency
< 10 kHz
Signal coupling
DC (AC/DC on the differential synchr. inputs)
Sampling rate, analog inputs
153.6 / 76.8 / 38.4 / 19.2 / 9.6 kHz
Frequency range
0...50 Hz to 0...50 kHz, sub-devided into 11 areas
Frequency resolution
400, 800, 1600, 3200, 6400, 12800 lines
Anti-aliasing
Dynamic adaptation
Envelope
Digital input filter, selectable
Measurement functions
Time waveform, spectrum, integration of the spectrum, envelope, orbit,
Overall values: shock pulse, acceleration (RMS), vibration velocity (peak, RMS)
Operation modes
Frequency band analysis, transient memory, online classification, trending
General parameters
Power supply
5.05V / 1.5A max. and 12V / 1.2A max.
Memory
RAM: 128 MB / Flash: 1000 MB
Temperature range, operation
- 20°C ... +60°C
Humidity
10% to 100%, dew permitted
Mechanical load
Shock:30 g / constant vibration: 2 g (10-150 Hz)
Protection class
IP 66 (EN 60529) / NEMA 4
Dimensions
approx. 260 x 130 x 150 mm (L x W x H)
Total weight
approx. 1.5 kg
Chapter 3: Technical data
3.1 Hardware
VIBRONET Signalmaster basic unit - VIB 5.802