Fluke 19xB-19xC-2x5C
Service Manual
5-14
Table 5-4. Input A&B Gain Calibration Points
Cal step
UUT input value (5500A NORMAL)
CL 0824
250 mV
CL 0799
Not for software versions V05.01 and V05.02
5
mV
CL 0800
12.5 mV
CL 0801
25 mV
CL 0802
50 mV
CL 0803
125 mV
CL 0805
500 mV
CL 0806
1.25 V
CL 0807
2.5 V
CL 0808
5 V
CL 0809
12.5 V
CL 0810
25 V
CL 0811
50 V (set 5500A to OPR!)
CL 0812
125 V
CL 0813
250 V
5.6.6 DMM Volt Gain
Warning
Dangerous voltages will be present on the calibration source
and connection cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to do the DMM Volt Gain calibration.
1.
Press
to select the first calibration step in Table 5-5.
2.
Connect the test tool to the 5500A as shown in Figure 5-7.