Performance Verification
4.8 Meter (DMM) Tests
4
4-29
4.8.2 Meter AC Voltage Accuracy & Frequency Response Test
Warning
Dangerous voltages will be present on the calibration source
and connecting cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to test the ac voltage measurement accuracy:
1.
Connect the test tool to the 5500A as for the previous test (see Figure 4-18).
2.
Select the following test tool setup:
•
Press
•
Press
to open the Measurement menu, and select
V ac
•
Press
to select MANUAL ranging; use
to select the ranges
3.
Set the range to the first test point in Table 4-13.
4.
Set the 5500A to source the appropriate ac voltage.
5.
Observe the reading and check to see if it is within the range shown under the
appropriate column.
6.
Continue through the test points.
7.
When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
Table 4-13. Meter Volts AC Measurement Verification Points
Range
5500A output V ac
Frequency
Meter Reading
500.0 mV
500.0 mV
60 Hz
494.0 to 506.0
1 kHz
486.0 to 514.0
10
kHz
>350.0
5.000 V
5.000 V
60 Hz
4.940 to 5.060
1 kHz
4.860 to 5.140
10
kHz
>3.500
50.00 V
50.00 V
60 Hz
49.40 to 50.60
1 kHz
48.60 to 51.40
10
kHz
>35.00
500.0 V
500.0 V
60 Hz
494.0 to 506.0
1 kHz
486.0 to 514.0
10
kHz
>350.0
1100 V (1.1 kV)
1000 V
60 Hz
0.980 to 1.020
1 kHz
0.960 to 1.040
10 kHz
> 0.700
Summary of Contents for ScopeMeter 19 B Series
Page 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Page 8: ......
Page 10: ......
Page 14: ...2 2 ...
Page 16: ......
Page 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Page 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Page 58: ......
Page 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Page 78: ......