Performance Verification
4.6 Scope Input A&B Tests
4
4-19
4.6.10 Input A AC Voltage Accuracy (HF) & Bandwidth Test
Proceed as follows to test the Input A high frequency automatic scope ac voltage
measurement accuracy, and the bandwidth:
1.
Connect the test tool to the 5500A as for the previous test (see Figure 4-5).
2.
Select the following test tool setup:
•
Recall the created SETUP 1 (see section 4.4.3): press
,
RECALL
,
select
SETUP 1
, press
RECALL SETUP
.
•
Press
, then press
READING 1
, and select
on A | V ac
.
•
Press
to select autoranging (
AUTO
in upper right LCD edge)
•
Using
and
change the sensitivity range to select manual sensitivity
ranging, and lock the Input A sensitivity range on 500 mV/div. (
AUTO
in upper
right LCD edge disappears)
3.
Set the 5500A to source a sine wave, to the first test point in Table 4-9.
4.
Observe the Input A reading and check to see if it is within the range shown under
the appropriate column.
5.
Continue through the test points.
6.
When you are finished, set the 5500A to Standby.
Table 4-9. HF AC Voltage Verification Points
UUT
5500A SC... MODE levsin
UUT
Model Voltage Frequency
Reading
A
all
2.545 Vpp
1 MHz
835 mV to 965 mV
all
2.545 Vpp
25 MHz
790 mV to 1.010 V
192B-C
2.545 Vpp
60 MHz
>630 mV
196B-C, 215C
2.545 Vpp
100 MHz
>630 mV
199B-C, 225C
2.545 Vpp
200 MHz
>630 mV
Summary of Contents for ScopeMeter 19 B Series
Page 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Page 8: ......
Page 10: ......
Page 14: ...2 2 ...
Page 16: ......
Page 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Page 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Page 58: ......
Page 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Page 78: ......