3. OPERATION
Page 77
© 2005-2007 DH Instruments, a Fluke Company
After the last log has been viewed, the option to clear the log,
<1no>
,
<2yes>
is
presented. Use
<2yes>
to remove all entries from the log. Use
<1no>
to
continue without altering the log.
To leave the log, use
[ESC]
.
3.5.8
<8CAL>
PURPOSE
To calibrate the RPM4-AD Hi and Lo Q-RPTs and adjust the on-board barometer. These
functions are considered part of RPM4-AD maintenance and are covered in the maintenance
section of this manual (see Sections 5.2, 5.3).
3.5.9
<9RESET>
PURPOSE
To reset various RPM4-AD settings to default or factory values.
PRINCIPLE
RPM4-AD stores its user definable settings in non-volatile memory. The reset menu allows the
user to selectively or completely reset these settings to factory defaults. This clears any settings
that the user has made, and should be used only to restore the RPM4-AD to a known state.
RPM4-AD goes through its power up sequence after any type of reset is executed.
RPM4-AD reset functions will change current settings to factory defaults. These may include
settings vital to RPM4-AD operation and affecting the calibration of the quartz reference
pressure transducers (Q-RPTs). Reset functions should only be used by qualified personnel
with knowledge of reset consequences. Reset functions should never be used “experimentally”.
OPERATION
To access the RESET menu, press
[SPECIAL]
,
<9reset>
. The display is:
1sets 2units 3ATest
4cal 5all
RESET menu choices include:
<1set>
to reset general system operating parameters (see Section 3.5.9.1).
<2units>
to reset unit of measure functions (see Section 3.5.9.2).
<3ATest>
not used in RPM4-AD models.
<4cal>
to reset internal calibration coefficients and modes (see Section 3.5.9.4).
<5all>
to reset all settings except ID and security password to factory default values
(see Section 3.5.9.5).