RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
© 2005-2007 DH Instruments, a Fluke Company
Page 78
3.5.9.1
<1SETS>
PURPOSE
Sets most general operating parameters back to default values. Does not affect
calibration coefficients or remote interfaces. The Reset – Sets resets are itemized
in Table 17.
Table 17.
Reset – Sets
RESET
RESULT
SEE SECTION
Active Q-RPT
Hi
3.2.1, 3.3.1
[UNIT]
Pressure unit of measure to kts for Hd Q-RPT and ft for Lo Q-
RPT
3.3.2
[MODE]
Measurement mode to ab differential..
3.3.3
[RANGE]
Hi Q-RPT in ab differential mode (Hd)
3.3.1
[HEAD]
0 cm height and Nitrogen medium
3.3
Stability Limit
3 Pa for A160K Q-RPT, 6 Pa for A350K Q-RPT
3.4.3
Upper Limit
102% of Hi Q-RPT range
3.4.4
Resolution
0.001 % FS of Hi Q-RPT default range
3.4.2
AutoZ
AutoZ ON for all Q-RPTs and measurement modes.
Z
offset
value not affected
3.5.1
AutoZ
P
atm,0
set to 101.325 kPa a
3.5.1, 3.2.2
Leak Check
15 second run time. Clear logged results
3.3.5
Screen Saver
10 minutes to activation
3.5.5.1
Key Sounds
Medium tone valid key press sound
3.5.5.2
Lo Vnt
Not used in RPM4-AD models.
3.5.7.4
Display RPT
display
3.3.6.4
ReadRate Automatic
3.5.7.2
3.5.9.2
<2 UNITS>
PURPOSE
Sets the six pressure units available under the UNIT function for various
measurement modes to the default selections (see Section 3.3.2, 3.5.6).
Sets the user defined unit to 1.000/Pa (see Section 3.5.6).
Sets the reference temperature for inWa unit to 20°C.
Sets the reference temperature for true airspeed units to 15°C.
3.5.9.3
<3AT
EST
>
This function is not used in RPM4-AD models.