3. OPERATION
Page 65
© 2005-2007 DH Instruments, a Fluke Company
inactive
TEST
port. When SDS is open, SDS temporary close can be used to
isolate and vent the specified Q-RPT.
Opening SDS with a pressure greater than Pmax! applied to the test port
may cause overpressure damage to the Q-RPT.
3.5.4.2
<2FULL TIME ON/OFF>
OPERATION
See
Section
3.2.8 for complete information on RPM4’s SDS (Self Defense
System) feature.
Turning SDS “full time OFF” eliminates normal SDS operation so that the
active TEST port remains open. However, the SDS overpressure function
(see Section 3.4.4.1) remains active.
To turn SDS full time ON/OFF, press
[SPECIAL]
,
<4SDS>
,
<2full time on/off >
.
The next display is:
The cursor is on the current state of SDS
ON/OFF. Select
<1on>
to set SDS for normal
operation and return to the MAIN RUN screen.
Select
<2off>
to open SDS permanently so
operation is as if SDS were not installed. Press
[ESC]
to return without making changes.
SDS full time:
1on 2off
If
<2off>
is selected and SDS is not already open, the SDS opening screen with
indication of the maximum pressure of the Q-RPT is displayed. This is the same
screen used when opening SDS using the
[SDS]
key (see Section 3.3.8). Check
that the pressure applied to the Q-RPT module
TEST
port does not exceed the
Pmax! value before confirming that SDS should be opened.
Opening SDS with a pressure greater than Pmax! applied to the TEST port
may cause overpressure damage to the Q-RPT.
With SDS “full time OFF”, when using parallel measurement, the active TEST
port is still switched between the Ps and Pt port depending upon
measurement mode (see Section 3.2.5).
3.5.5
<5PREFS>
PURPOSE
To access a menu of RPM4-AD operational preferences and functions.
OPERATION
To access the PREFS menu press
[SPECIAL]
,
<5prefs>
. The display is:
1ScrSvr 2sound 3time
4ID 5level