RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
© 2005-2007 DH Instruments, a Fluke Company
Page 104
SS
n
Purpose
Read or set the current pressure stability limit.
Command
Query
“SS
limit
”
“SS?”
Classic “SS=
limit
”
“SS”
Optional Suffix
“n”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
Arguments
Limit:
The stability limit in the current pressure unit of measure.
Remarks
The stability limit can be read and set as a pressure. The stability limit is used as the
Ready/Not Ready
criteria. It the Q-RPT is in an altitude or airspeed unit, then the stability
setting is represented in the pressure unit “Pa”.
Example
(enhanced)
Cmd sent:
Query reply:
“SS .1”
“0.10 kPa/s”
(No reply from GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“SS? .1”
“0.10 kPa/s”
Example
(classic)
Sent:
Reply:
“SS=.1”
“0.10 kPa/s”
Errors ERR#
6:
ERR# 10:
The argument was invalid.
Invalid or missing command suffix.
See Also
“SS%”
3.4.3
TIME
Purpose
Read or set the RPM4-AD internal clock.
Command
Query
“TIME
hh:mmXX
”
“TIME?”
Classic “TIME=
hh:mmXX
”
“TIME”
Arguments
hh:mm:
XX:
The time in a 12 hour format using a colon delimiter
”am” or “pm”
Example
(enhanced)
Cmd sent:
Query reply:
“TIME 12:52PM”
“12:52pm”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“TIME? 12:52PM”
“12:52pm”
Example
(classic)
Sent:
Reply:
“TIME=12:52PM”
“12:52pm”
Errors
ERR# 7:
Missing or improper program message argument(s).
See Also
“DATE”
3.5.5.3
UCOEF
n
Purpose
To identify the coefficient used to convert 1 Pascal to the current pressure unit of measure.
Command
Query
“UCOEF
n Pres
”
“UCOEF
n Pres, Temp
”
“UCOEF
n
?”
Classic “UCOEF
n = Pres, Temp
”
“UCOEF
n
”
Arguments
Pres:
Temp:
Pressure in Pa to convert to the Q-RPT’s measurement units.
Optional reference temperature in
°
C if the Q-RPT unit is a “True” airspeed.
Optional Suffix
“n”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
Remarks
The RPM4-AD handles all pressure values internally in Pascal. The coefficient replied is
equivalent of
Pres
Pa in the current pressure unit of measure. If the optional argument
Pres
is not given, then 1Pa is assumed and converted. The
Temp
argument is only valid for True
airspeed units.
Example
(enhanced)
Query sent:
Query reply:
“UCOEF?”
(current units are kPa)
“0.0010000000”
Example
(classic)
Sent:
Reply:
“UCOEF=100” (current units are kPa)
“0.1000000000”
See Also
3.3.2, 3.5.6, 7.2