Innovative High Temperature Fiber Optic Infrared Pyrometers
Maintenance
85
Figure 59: Removing the Fiber optic Cable from the Optical Head
11.4.1.2. Removing the Fiber Optic Cable from the Electronics Housing
Complete the following steps to disconnect the fiber optic cable from the electronics housing:
1. First loosen the cable connecting sleeve.
2. Loosen the cable receptacle screw to release the cable.
3. Pull cable from electronics enclosure, and immediately place a protective cap over the
end of the fiber optic cable. Do not use adhesive tape on the cable end.
Figure 60: Removing the Fiber optic Cable from the Electronics Housing
Turn 1.3 mm (0.050“) hex
wrench counter clockwise
until cable is loose
Put cable out
First loosen cable
connecting sleeve
Then loosen screw and pull
cable from coupling.