Endurance® Series
Users Manual
82
Table 11: Fail-safe Error Codes
Condition
2-Color
1-Color
(wide band)**
1-Color*
(narrow band)**
Priority
Heater control temperature over range
ECHH
ECHH
ECHH
1 (high)
Heater control temperature under range
ECUU
ECUU
ECUU
2
Internal temperature over range
EIHH
EIHH
EIHH
3
Internal temperature under range
EIUU
EIUU
EIUU
4
Wide band detector failure
EHHH
EHHH
<temperature>
5
Narrow band detector failure
EHHH
<temperature> EHHH
6
Energy too low
EUUU
<temperature> <temperature>
7
Attenuation too high (>98%)
EAAA
<temperature> <temperature>
8
Attenuation too high >95%
("dirty lens", relay will go to “alarm” state)
<temperature> <temperature> <temperature>
9
2-color temperature under range
EUUU
<temperature> <temperature>
10
2-color temperature over range
EHHH
<temperature> <temperature>
11
1-color temperature (wide) under range
<temperature> EUUU
<temperature>
12
1-color temperature (wide) over range
<temperature> EHHH
<temperature>
13
1-color temperature (narrow) under range <temperature> <temperature> EUUU
14
1-color temperature (narrow) over range
<temperature> <temperature> EHHH
15 (low)
* only available via RS485 or LAN / Ethernet command
** Wide and narrow band stands for the first and the second wavelength in 2-Color mode
*** Note that the activation levels for these conditions may be set to different values.
(e.g., "dirty lens" at 95%, EAAA at 98%)
11.2.2. Analog Output current values in dependence of Fail-Safe Error Codes
The relay is controlled by the temperature selected on the display. If any failsafe code appears
on the display, the relay changes
to the “abnormal” state. The exception is the “dirty window”
condition. This causes the relay to change state, leaving a normal numerical temperature
output.
Table 12: Current Output Values in accordance to an Error
Error Code 0
– 20 mA Output
4
– 20 mA Output
no error
according to temperature according to temperature
ECHH
21 to 24 mA
21 to 24 mA
ECUU
0 mA
2 to 3 mA
EIHH
21 to 24 mA
21 to 24 mA
EIUU
0 mA
2 to 3 mA
EUUU
0 mA
2 to 3 mA
EHHH
21 to 24 mA
21 to 24 mA
EAAA
0 mA
2 to 3 mA
If two or more errors occur simultaneously, the error with the highest priority overrules the lower
priority errors. The highest priority error will be displayed on the 7-segment temperature display
and the assigned analog output (current) value (see Table 12) will be set. For instance, in 2-
Color mode, if the internal ambient temperature is over the limit and the attenuation is to high
too, the unit outputs EIHH to the temperature display and sets an analog output current of 21
mA on the analog current loop output lines. However, since 2-Color wide band and narrow band
temperatures may all be presented simultaneously through RS485 or LAN / Ethernet interface,
their over and under range conditions are independent.