700G Series
Users Manual
10
RS-232 Interface
The Product includes an RS-232 interface. You can use serial communication to configure and calibrate
the Product and move measurement data from the Product to a PC. An RS-232/USB cable is sold
separately and includes 700G/TRACK Software. For specifications on the interface, see
Specifications
.
W
Warning
To prevent possible fire, explosion, or personal injury, do not use the RS-232 interface in
hazardous areas. The RS-232 interface port must be sealed when used in a hazardous area.
Specifications
Available Input Ranges
................................See Ranges and Resolution tables for available ranges in psi plus equivalent
ranges and resolution for all pressure units.
Accuracy 700G Ranges
Positive Pressure (700G01, 700G02).........±0.1 % Full Scale
Positive Pressure (all other ranges)............±0.05 % Full Scale
Vacuum.......................................................±0.1 % Full Scale
Temperature Compensation .......................15 °C to 35 °C (59 °F to 95 °F) to rated accuracy
Note
For temperatures from -10 °C to +15 °C and +35 °C to +55 °C, add 0.003 % FS / °C
.
Accuracy 700RG Ranges
Positive Pressure........................................±0.04 % rdg ±0.01 % Full Scale
Vacuum (700RG05)....................................±0.05 % Full Scale
Vacuum (all other ranges) ..........................±0.1 % Full Scale
Temperature Compensation .......................0 °C to +50 °C (+32 °F to +122 °F) to rated accuracy
Note
For temperatures from -10 °C to 0 °C and +50 °C to +55 °C, add 0.005 % FS / °C.
Media Compatibility
700G01, 700G02, 700G04,
700G05, 700RG05......................................any clean, dry, non-flammable, non-corrosive gas
All other ranges
100 psi to 1000 psi ..................................any non-flammable liquids or gases compatible with 316 stainless steel
Above 1000 psi .......................................any non-flammable, non-toxic, non-oxidizing liquid or gas compatible with 316
stainless steel
Mechanical Specifications
Dimensions .................................................11.4 cm x 12.7 cm x 3.7 cm (4.5 in x 5.0 in x 1.5 in)
Pressure
Connection ..............................................¼ in NPT Male
Housing ...................................................Cast ZNAL
Display
...........................................................5-1/2 Digits, 16.5 mm (0.65 in) high
20-Segment bar graph, 0 % to 100 %
Power
Battery ........................................................3 x AA alkaline IEC LR6
Battery Life..................................................1500 hours without backlight (continuous on), 2000 hours at slow sample rate
Summary of Contents for 700G0