Pressure Gauges
11
Environmental
Temperature
Operating ................................................-10 °C to +55 °C (+14 °F to +131 °F)
Storage
With Batteries ......................................Per battery manufacturer’s specification, not to exceed storage specification
without batteries.
Without Batteries .................................-40 °C to +70 °C (-40 °F to +158 °F)
Altitude
Operating ................................................2000 m
Storage ...................................................12 000 m
Relative Humidity........................................10 % to 95 % non-condensing
Safety .........................................................IEC 60079
-
0, IEC 60079
-
11, IEC 61010-1: Pollution Degree 2
Ingress Protection Rating ...........................IEC 60529: IP67 (with battery door seal and serial-port plug installed)
Electromagnetic Compatibility (EMC
)
International................................................IEC 61326-1: Portable Electromagnetic Environment; IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-
voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to
radio reception in such environments.
Korea (KCC) ...............................................Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) .................................................47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103.
RS-232 Interface
...........................................The Serial Interface (J4) is a 3.5 mm Stereo Jack connector. Use the RS-232/
USB cable sold with the 700G/TRACK Software. USB to RS-232 converter with
5 V RS-232 signals. Connection to RS-232 Interface in non-hazardous area only:
Ui = 18 V, Pi = 0.5 W.
Compliance Marks
)
............................................................Class I, Division 2, Groups A-D
P
.................................................................................
(
II 3 G Ex ic IIB T6 Gc
SIRA 17ATEX4160X
Input Entity Parameters:
Ta = -10 °C...+55 °C
Ui = 18 V, Pi = 0.5 W (NON HAZARDOUS AREA)
Summary of Contents for 700G0
Page 4: ...700G Series Users Manual ii ...
Page 18: ...700G Series Users Manual 14 ...