5820A
Service Manual
3-32
•
BNC(f) to 3.5 mm(m) adapter (2)
•
N to BNC cable supplied with the Scope Calibrator
•
second N to BNC cable
Connect the N to BNC cable supplied with the Scope Calibrator to the Calibrator
Mainframe’s CHAN 1 connector. Connect the other end of the N to BNC cable to one
BNC(f) to 3.5 mm(m) adapter then to the DSO’s sampling head through the 3 dB
attenuator. Refer to Figure 3-6 for the proper equipment connections.
Using the second BNC(f) to 3.5 mm(m) adapter and N to BNC cable, connect the
Calibrator Mainframe’s EXT TRIG (channel 5) connector to the 11801’s Trigger Input.
CHAN 2
CHAN 4
EXT TRIG
EXT TRIG
SOURCE/MEASURE
CHAN 1
CHAN 2
CHAN 3
CHAN 4
CHAN 5
20V PK
MAX
CHAN 1-5
AUX
INPUT
20V PK
MAX
20V PK
MAX
CAT
130V
PK
MAX
SOURCE
30V DC
MAX
MEASURE
5820A
OSCILLOSCOPE
CALIBRATOR • 2GHz
5820A
BNC(F) to
3.5 mm (m)
Adapter
3 dB Attenaator
3.5 mm (m/f)
Tek 11801
With 5D26 Sampling Head
5820A
Cable
yu058f.eps
Figure 3-6. Setup for Edge Rise Time Verification
The Calibrator Mainframe should have the Edge menu on the display. Press
on the
Calibrator Mainframe to activate the output. Press the softkey under TRIG to select the
TRIG/1 External Trigger output. Program the Calibrator Mainframe to output
250 mV @ 1 kHz. Set the DSO to these parameters:
Digital Storage Oscilloscope Setup
Main Time Base position (initial)
40 ns
Horizontal scale
500 ps/div
Measurement Function
Rise Time
1.
Program the Calibrator Mainframe to output the voltage and frequency listed in
Table 3-16. Press
on the Calibrator Mainframe to activate the output.
2.
Change the vertical scale of the DSO to the value listed in the table. Adjust the main
time base position and vertical offset until the edge signal is centered on the display.
Record the rise time measurement in column A of Table 3-16.
3.
Correct the rise time measurement by accounting for the SD-22/26 sampling head’s
rise time. The SD-22/26 rise time is specified as <28 ps. Column B = sqrt((Column
A)
2
- (SD-22/26 rise time)
2
).
4.
The edge rise time measured should be less than the time indicated in Table 3-16.
Summary of Contents for 5820A
Page 12: ...5820A Service Manual viii ...
Page 14: ...5820A Service Manual 1 2 ...
Page 28: ...5820A Service Manual 1 16 ...
Page 30: ...5820A Service Manual 2 2 ...
Page 38: ...5820A Service Manual 2 10 ...
Page 102: ...5820A Service Manual 3 64 ...
Page 104: ...5820A Service Manual 4 2 ...
Page 110: ...5820A Service Manual 5 2 ...
Page 140: ...5820A Service Manual 6 2 ...
Page 156: ...5820A Service Manual 6 18 ...