Options
Verification Tables
5
5-11
2.
Program the Calibrator Mainframe to 3.5 V p-p at each frequency listed in Table
5-6. Press
on the Calibrator Mainframe to activate the output.
3.
Set HP 8592L start frequency to the Calibrator Mainframe output frequency. Set HP
8592L stop frequency to 10 times the Calibrator Mainframe output frequency. Set
the HP 8592L reference level at +19 dBm.
4.
Record the harmonic level reading for each frequency and harmonic listed in Table
5-6. For harmonics 3, 4, and 5, record the highest harmonic level of the three
measured. Harmonics should be below the levels listed in the Deviation column of
Table 5-6.
5-12. Verification Tables
Table 5-4 through Table 5-7 are to be used to verify channels 2-5. The verification test
points are provided here as a guide when verification to one-year specifications is
desired. For more information on verification, see Chapter 3.
Table 5-4. GHz Leveled Sine Wave Frequency Verification (Channels 1, 2, and 5)
Calibrator Mainframe
Frequency
PM 6680
Settings
PM 6680
Reading
Tolerance
(Output @ 3.5 V p-p)
Channel
Filter
(Frequency)
±
(Hz)
10 MHz
A
Off
3.3
601 MHz
C
Off
198.3
1000 MHz
C
Off
330
1500 MHz
C
Off
495
2000 MHz
C
Off
660
2100 MHz
C
Off
693
Table 5-5. Fast Edge Rise Time for Channels 1, 2, and 5
Channel
Function
Measurement
Amplitude
Frequency
Measured
Value
Deviation
Spec (s)
1-5
edge
rise time
0.25
1000000
150 ps
Summary of Contents for 5820A
Page 12: ...5820A Service Manual viii ...
Page 14: ...5820A Service Manual 1 2 ...
Page 28: ...5820A Service Manual 1 16 ...
Page 30: ...5820A Service Manual 2 2 ...
Page 38: ...5820A Service Manual 2 10 ...
Page 102: ...5820A Service Manual 3 64 ...
Page 104: ...5820A Service Manual 4 2 ...
Page 110: ...5820A Service Manual 5 2 ...
Page 140: ...5820A Service Manual 6 2 ...
Page 156: ...5820A Service Manual 6 18 ...