5820A
Service Manual
3-22
Table 3-5. DC Voltage Verification (cont.)
Nominal Value
(V dc)
Measured Value
(V dc)
Deviation
(V dc)
1-Year Spec.
0.5
150
µ
V
-0.5
150
µ
V
1.35
362.5
µ
V
-1.35
362.5
µ
V
2.19
572
µ
V
-2.19
572
µ
V
2.2
572
µ
V
-2.2
572
µ
V
6.6
1.67 mV
-6.6
1.67 mV
10.99
2.77 mV
-10.99
2.77 mV
11
2.77 mV
-11
2.77 mV
70.5
17.65 mV
-70.5
17.65 mV
130
32.5 mV
-130
32.5 mV
6.599 (50
Ω
)
16.5 mV
Table 3-6. DC Voltage Verification at 50
Ω
Calibrator
Mainframe
Output
HP 3458A Rdg (V dc)
Reading x Correction
Tolerance (V dc)
0 mV
0.00004 V
2.49 mV
4.623E-05 V
-2.49 mV
4.623E-05 V
9.90 mV
6.475E-05 V
-9.90 mV
6.475E-05 V
24.9 mV
0.0001023 V
-24.9 mV
0.0001023 V
109.9 mV
0.0003148 V
-109.9 mV
0.0003148 V
499 mV
0.0012875 V
-499 mV
0.0012875 V
2.19 V
0.005515 V
-2.19 V
0.005515 V
6.599 V
0.0165375 V
-6.599 V
0.0165375 V