Figure 3-5. Setup for AC Voltage Frequency Verification
Set the Calibrator Mainframe to the Volt menu. Press
on the Calibrator Mainframe
to activate the output. Then follow these steps to verify ac voltage frequency.
1.
Set the PM 6680’s FUNCTION to measure frequency on channel A with auto
trigger, measurement time set to 1 second or longer, 1M
Ω
impedance, and filter off.
2.
Using the N to BNC cable, connect the CHAN 1 connector on the Calibrator
Mainframe to PM 6680 channel A.
3.
Program the Calibrator Mainframe to output 2.1 V at each frequency listed in Table
3-9.
4.
Allow the PM 6680 reading to stabilize, then record the PM 6680 reading for each
frequency listed in Table 3-9. Compare to the tolerance column of Table 3-9.
Table 3-9. AC Voltage Frequency Verification
Calibrator Mainframe
Frequency
(Output @ 2.1 V p-p)
PM 6680 Reading
(Frequency)
Tolerance
10 Hz
3.3 E
-6
100 Hz
33 E
-6
1 kHz
330 E
-6
10 kHz
3,300 E
-6
3-32. DC Measurement Verification
This procedure uses the following equipment:
•
Fluke 5520A Calibrator
1.
Select Voltage Measurement mode.
Summary of Contents for 5820A
Page 12: ...5820A Service Manual viii ...
Page 14: ...5820A Service Manual 1 2 ...
Page 28: ...5820A Service Manual 1 16 ...
Page 30: ...5820A Service Manual 2 2 ...
Page 38: ...5820A Service Manual 2 10 ...
Page 102: ...5820A Service Manual 3 64 ...
Page 104: ...5820A Service Manual 4 2 ...
Page 110: ...5820A Service Manual 5 2 ...
Page 140: ...5820A Service Manual 6 2 ...
Page 156: ...5820A Service Manual 6 18 ...