Calibration and Verification
Verification
3
3-25
Table 3-8. AC Voltage Verification at 50
Ω
Calibrator
Mainframe
Output
(1 kHz)
HP 3458A
Range
Topline
Reading
Baseline
Reading
Peak-to-Peak
Peak-to-Peak x
Correction
Tolerance
(
±
V)
1 mV
100 mV dc
0.000043
-1 mV
100 mV dc
0.000043
10 mV
100 mV dc
0.000065
-10 mV
100 mV dc
0.000065
25 mV
100 mV dc
0.000103
-25 mV
100 mV dc
0.000103
110 mV
100 mV dc
0.000315
-110 mV
100 mV dc
0.000315
500 mV
1 V dc
0.00129
-500 mV
1 V dc
0.00129
2.2 V
10 V dc
0.00554
-2.2 V
10 V dc
0.00554
6.6 V
10 V dc
0.01654
-6.6 V
10 V dc
0.01654
3-31. AC Voltage Frequency Verification
This procedure uses the following equipment:
•
PM 6680 Frequency Counter with an ovenized timebase (Option PM 9690 or PM
9691)
•
N to BNC cable supplied with the Scope Calibrator
Refer to Figure 3-5 for the proper equipment connections.