Introduction and Specifications
General Specifications
1
1-7
Temperature Coefficient
Temperature Coefficient for temperatures outside tcal
±
5
°
C is 0.1X/
°
C of
1-year specification.
Relative Humidity
Operating: < 80 % to 30
°
C, < 70 % to 40
°
C,< 40 % to 50
°
C
Storage: < 95 %, noncondensing
Altitude
Operating: 3,050 m (10,000 ft) maximum
Nonoperating: 12,200 m (40,000 ft) maximum
Safety
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Interface Impedance
The 5820A is designed to drive both 50
Ω
and 1M
Ω
loads.
Analog Low Isolation
20 V
EMC
Complies with EN 61326-1
Line Power
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 to 63 Hz
Line Voltage Variation:
±
10 % about line voltage setting
Power Consumption
250 VA
Dimensions
Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm
(0.6 inch) for feet on bottom of unit;
Width: 43.2 cm (17 inches), standard rack width
Depth: 47.3 cm (18.6 inches) overall.
Weight
20 kg (44 pounds)
Summary of Contents for 5820A
Page 12: ...5820A Service Manual viii ...
Page 14: ...5820A Service Manual 1 2 ...
Page 28: ...5820A Service Manual 1 16 ...
Page 30: ...5820A Service Manual 2 2 ...
Page 38: ...5820A Service Manual 2 10 ...
Page 102: ...5820A Service Manual 3 64 ...
Page 104: ...5820A Service Manual 4 2 ...
Page 110: ...5820A Service Manual 5 2 ...
Page 140: ...5820A Service Manual 6 2 ...
Page 156: ...5820A Service Manual 6 18 ...