Calibration and Verification
Verification
3
3-27
2.
Using the N to BNC cable, connect the CHAN 1 connector on the Calibrator
Mainframe to the Fluke 5520A Calibrator voltage binding posts using a BNC to
banana connector.
Record the 5820A reading for each voltage in Table 3-10.
Table 3-10. DC Voltage Measurement Verification
Nominal Value
(V dc)
Measured Value
Deviation
1-year Spec.
(V p-p)
0 V
0.001
1 V
0.0015
2.5 V
0.00225
5 V
0.0035
6.5 V
0.02625
9.9 V
0.03475
-1 V
0.0015
-2.5 V
0.00225
-5 V
0.0035
-6.5 V
0.02625
-9.9 V
0.03475
3-33. Current Verification
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter with cable
•
1
Ω,
10
Ω,
100
Ω,
and
1000
Ω
precision resistor assemblies
Remove the front panel Current Loop using the 2 mm hex head driver.
3-34. DC Current Verification
1.
Connect the DMM directly to the vacant banana jacks on the calibrator.
2.
Program the Calibrator Mainframe to output the Currents listed in Table 3-11. Press
on the Calibrator Mainframe to activate the output.
3.
Allow the HP 3458A reading to stabilize, then record the HP 3458A reading for each
current in Table 3-11.
4.
Compare result to the tolerance column.
Summary of Contents for 5820A
Page 12: ...5820A Service Manual viii ...
Page 14: ...5820A Service Manual 1 2 ...
Page 28: ...5820A Service Manual 1 16 ...
Page 30: ...5820A Service Manual 2 2 ...
Page 38: ...5820A Service Manual 2 10 ...
Page 102: ...5820A Service Manual 3 64 ...
Page 104: ...5820A Service Manual 4 2 ...
Page 110: ...5820A Service Manual 5 2 ...
Page 140: ...5820A Service Manual 6 2 ...
Page 156: ...5820A Service Manual 6 18 ...