Front Panel Operation
Techniques for Reducing Transfer Errors
4
4-27
4.
The choices available under the "Adjust the" softkey are DAY, MONTH, YEAR,
HOUR, MINUTE, and SECOND. The softkeys autorepeat if you hold them down.
Press the [^] key to increment, or the [v] key to decrement the selected parameter.
Note
Time is shown in 24-hour format and includes leading zeros. The format is hh:mm:ss
where hh = hours, mm = minutes, and ss = seconds.
5.
Return the rear panel CALIBRATION STORE switch to NORMAL.
6.
To clear the clock setting display, press the "DONE Setting Clock" softkey followed
by "DONE With Utils."
4-35.
[SPEC] Key
At any time during front panel operation, you can display the uncertainty specification of
the present output of the 5790A by pressing [SPEC]. To return to normal operation, press
the [SPEC] key again. An example of the specification display follows:
add08y.eps
Note
For INPUT 1 OR 2, the specification displayed is the measurement mode
absolute uncertainty specification. For WIDEBAND, the specification
displayed is the total uncertainty including flatness. For the SHUNT input,
no specification is displayed. The days since calibration relates to the
calibration procedure for the active input (Periodic or WIDEBAND).
4-36.
Techniques for Reducing Transfer Errors
In making ac-dc transfers for the purpose of determining absolute ac voltage, errors can
come from dc reversal, loading (both cable induced and instrument input), thermal emfs,
changes in mechanical contact, spurious ground currents in the signal leads, and EMI.
Knowing more about these errors can help reduce the amount of time you spend
experimenting and debugging to get the best possible results. Once you are confident that
error sources are minimized, a good practice is to take three measurements. That way, if
one measurement is faulty, it stands out.
4-37.
DC Reversal
In measurement mode, the 5790A cancels potential dc reversal errors by chopping its
internal dc reference into a low-frequency square wave. In transfer mode, it is up to you
to correct for dc reversal to meet specifications. Apply one polarity of dc to the transfer
standard INPUT and press the "Ref" softkey, then reverse the input polarity and press the
"Ref Avg" softkey. In this way you establish a reference that consists of the average of
both polarities of the dc input.
Summary of Contents for 5790A
Page 4: ......
Page 12: ...5790A Operators Manual vi...
Page 14: ...5790A Operators Manual viii...
Page 16: ...5790A Operators Manual 1 2...
Page 34: ...5790A Operators Manual 2 2...
Page 40: ...5790A Operators Manual 2 8...
Page 42: ...5790A Operators Manual 3 2...
Page 48: ...5790A Operators Manual 3 8...
Page 108: ...5790A Operators Manual 6 2...
Page 146: ...5790A Operators Manual 7 2...
Page 164: ...5790A Operators Manual 7 20...
Page 166: ...5790A Operators Manual 8 2...
Page 170: ...5790A Operators Manual 8 6...
Page 172: ...5790A Operators Manual 9 2...
Page 176: ...5790A Operators Manual...
Page 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 186: ...5790A Operators Manual B 2...
Page 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Page 188: ...5790A Operators Manual B 4...
Page 204: ...5790A Operators Manual C 16...
Page 210: ...5790A Operators Manual D 6...