5790A
Operators Manual
4-10
5790A
AC MEASUREMENT
STANDARD
5790A
EXT GRD
UUT SOURCE
1
2
Use HEAVY BRAID
3
INPUT
PLANE OF
REFERENCE
GUARD
add10f.eps
Figure 4-2. Recommended Test Lead Connections to Test a Source
4-13.
External Guard Connection
For some conditions, such as low-frequency common mode interference, better results
may be obtained by using the external guard configuration. Select external guard by
pressing [EX GRD] so that the keycap indicator is lit. This disconnects the LO from
GUARD and connects the internal guard to the GUARD binding post. A connection from
the GUARD binding post can then be made to the common earth ground point for the
system.