C-1
Appendix C
Calibration Constant Information
The constants in these tables are arranged by group. Each group is stored as a block in
nonvolatile memory. The value given for each constant in this list is the default assigned
before the instrument is first calibrated. Defaults are reinstated if you perform a format of
the EEPROM ALL or CAL areas.
Group ZC_BASIC: Internally Calibrated DAC, Sensor, and A/D Parameters
Name Default
Function
DAC_Z1
398.0
Reference DAC zero, coarse channel
DAC-Z2
17500.00
Reference DAC zero, fine channel
DAC_RATIO
16500.0
Reference DAC coarse/fine gain ratio
AD_X1_Z
0.0
A/D x1 range zero
AD_X1_G
-7.23E-6
A/D x1 range gain
AD_X10_Z
0.0
A/D x100 range zero
AD_X10_G
7.23E-8
A/D x100 range gain
NULLDAC_Z
0.0
Null DAC zero
NULLDAC_G
6560.0
Null DAC gain
SENSOR_C1
1.0
Main thrmal sensor linearization
SENSOR_C2
0.0
Main thrmal sensor linearization
OF_VSQ
0.0
Main thrmal sensor turnover correction
Summary of Contents for 5790A
Page 4: ......
Page 12: ...5790A Operators Manual vi...
Page 14: ...5790A Operators Manual viii...
Page 16: ...5790A Operators Manual 1 2...
Page 34: ...5790A Operators Manual 2 2...
Page 40: ...5790A Operators Manual 2 8...
Page 42: ...5790A Operators Manual 3 2...
Page 48: ...5790A Operators Manual 3 8...
Page 108: ...5790A Operators Manual 6 2...
Page 146: ...5790A Operators Manual 7 2...
Page 164: ...5790A Operators Manual 7 20...
Page 166: ...5790A Operators Manual 8 2...
Page 170: ...5790A Operators Manual 8 6...
Page 172: ...5790A Operators Manual 9 2...
Page 176: ...5790A Operators Manual...
Page 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 186: ...5790A Operators Manual B 2...
Page 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Page 188: ...5790A Operators Manual B 4...
Page 204: ...5790A Operators Manual C 16...
Page 210: ...5790A Operators Manual D 6...