2635A
Users Manual
7-6
Table 7-2. Recommended Test Equipment
Instrument Type
Minimum Specification
Recommended
Model
Multifunction Calibrator
DC Voltage:
Fluke 5700A
Range: 90 mV to 300V dc
Accuracy: .005%
AC Voltage:
Frequency
Voltage
Accuracy
1 kHz
29 mV to 300V
0.05%
100 kHz
15 mV to 300V
1.25%
Frequency:
10 kHz
1V rms
.0125%
Decade Resistance
Source
Ohms
Accuracy
General Resistance
Inc. Model RDS 66A
290
Ω
0.0125%
2.9 k
Ω
0.0125%
29 k
Ω
0.0125%
290 k
Ω
0.0125%
2.9 M
Ω
0.0125%
Note
The 5700A Calibrator provides 0.05% accuracy
(rated) on the 3.0 k
Ω
, 30 k
Ω
, 300 k
e
and 3.0
M
Ω
ranges. The 5700A can be used for 0.06%
accuracy on the 300
Ω
range.
Mercury Thermometer
0.02 degrees Celsius resolution
Princo ASTM-56C
Thermocouple Probe
Type T
Fluke P-20T
Room Temperature
Oil/Water Bath
Thermos bottle and cap
Multimeter
Me5V dc
Fluke 77
Signal Generator
Sinewave, 0.5 to 1V rms 10 Hz to 5 kHz
Fluke 6011A
Alternate Equipment List
(Minimum specifications are the same as in the Standard Equipment List)
Instrument Type
Recommended Model
DC Voltage Calibrator
Fluke 5700A
DMM Calibrator
Fluke 5100B (for AC Volts only)
Function/Signal Generator
Philips PM5193 or Fluke 6011A
Decade Resistance Source
Gen Rad 1433H
Summary of Contents for 2635A
Page 8: ...2635A Users Manual vi...
Page 26: ...2635A Users Manual xxiv...
Page 29: ...Preparation for Use Introduction 1 1 3...
Page 50: ...2635A Users Manual 1 24...
Page 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Page 88: ...2635A Users Manual 2 38...
Page 180: ...2635A Users Manual 5 10...
Page 212: ......
Page 234: ...2635A Users Manual B 4...
Page 250: ...2635A Users Manual D 8...
Page 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Page 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Page 268: ...2635A Users Manual 4...