Appendices
Specifications
A
A-17
Relative Humidity (Non-Condensing)
90% maximum for 0°C to 28°C
75% maximum for 28°C to 35°C
50% maximum for 35°C to 50°C
35% maximum for 50°C to 60°C
(Except 70% maximum for 0°C to 35°C, 0% maximum for 40°C to 50°C, and 20%
maximum for 50°C to 60°C for the 300 k
e
, 3 M
e
, and 10 M
e
ranges.)
Altitude
Operating: 2,000 m maximum
Non-operating: 12,200m maximum
Vibration
0.7g at 15 Hz
1.3g at 25 Hz
3g at 55 Hz
Shock
30g half sine per Mil-T-28800
Bench handling per Mil-T-28800
General
The general specifications are provided in the following paragraphs.
Channel Capacity
21 Analog Inputs
4 Alarm Outputs
8 Digital I/O (inputs/outputs)
Measurement Speed
Slow rate: 4 readings/second nominal
Fast rate: 17 readings/second nominal
1.5 readings/second nominal for ACV and high-
e
inputs
For additional information, refer to Typical Scanning Rated and Maximum Autoranging
Time.
Nonvolatile Memory Life
>10 unpowered instrument years for 0°C to 28°C
>3 unpowered instrument years for 0°C to 50°C
>2 unpowered instrument years for 50°C to 70°C
Stores: Real-time clock, set-up configuration, and measurement data.
Common Mode Voltage
300V dc or ac rms maximum from any analog input (channel) to earth provided that
channel to channel maximum voltage ratings are observed.
Summary of Contents for 2635A
Page 8: ...2635A Users Manual vi...
Page 26: ...2635A Users Manual xxiv...
Page 29: ...Preparation for Use Introduction 1 1 3...
Page 50: ...2635A Users Manual 1 24...
Page 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Page 88: ...2635A Users Manual 2 38...
Page 180: ...2635A Users Manual 5 10...
Page 212: ......
Page 234: ...2635A Users Manual B 4...
Page 250: ...2635A Users Manual D 8...
Page 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Page 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Page 268: ...2635A Users Manual 4...