Electrical
Installation
Tester
Memory
Mode
37
Memory Mode
(Model 1653B and 1654B Only
)
You can store measurements on the tester:
•
1653B – up to 444
•
1654B – up to 1500
The information stored for each measurement consists of the test function and
all user selectable test conditions.
Data for each measurement is assigned a data set number, data subset
number, and a data id number. Memory location fields are used as described
below.
Field Description
Use the data set field (a) to indicate a location such as a
room or electrical panel number.
Use the data subset field (b) for circuit number.
The data id number field (c) is the measurement number.
The measurement number automatically increments. The
measurement number can also be set to a previously used
value to overwrite an existing measurement.
To enter Memory mode:
1. Press
the
to enter Memory mode.
The display changes to a memory mode display. In Memory mode, the
icon appears on the display.
1653B:
The primary numeric display will be active with the left two digits
(a) indicating the data set number (1-99) and the right two digits (b)
indicating the data subset number. The decimal point separating these two
values will be active. The secondary numeric display (c) will be active
indicating the data id number (1-444). One of the memory locations, a, b,
or c, will flash to indicate that you can change the number using the arrow
keys
.
1654B:
The primary numeric display shows the data set number (a,
1-9999). The secondary numeric display shows the data subset number (b,
1-9999). The data id number (c, 1-9999) appears after you press
several times. One of the memory locations, a, b, or c, will flash to indicate
that you can change the number using the arrow keys
.