RPM4™ OPERATION AND MAINTENANCE MANUAL
Page
80
Table 12.
Reset – Cal
RESET
RESULT
SEE SECTION
All Q-RPT Calibration
Coefficients
PA to zero, PM to 1
5.2.1.1, 5.2.7
Axxx Q-RPT absolute and
negative gauge mode
ON 5.2.5
On-board Barometer
Calibration Coefficients
PA to zero, PM to 1
5.3
Calibration Date
Set all dates to 19800101
5.2.7, 5.3
AutoZ values
Axxx Q-RPTs: All P
offset
values to zero for
absolute mode and 101325 Pa for gauge and
negative gauge modes
Gxxx, BGxxx Q-RPTs: All P
offset
values to zero
3.5.1
AutoZ function
ON, all Q-RPTs, all measurement modes
3.5.1
3.5.9.5
<5 ALL>
PURPOSE
The reset - all function clears and deletes large amounts of user defined
information including critical calibration data.
Combines all resets in one global reset command that clears the entire user
section of non-volatile memory except the ID function (see Section 3.5.5.4) and
the security level password (see Section 3.5.5.5) returning RPM4 to the
“as delivered” condition. The Reset – All resets are itemized in Table 13.
Table 13.
Reset – All
RESET
RESULT
SEE SECTION
Reset – Sets
All the resets of Reset - Sets
3.5.9.1
Reset – Units
All the resets of Reset - Units
3.5.9.2
Reset – Atest
All the resets of Reset - ATest
3.5.9.3
Reset – Cal
All the resets of Reset - Cal
3.5.9.4
Remote Interfaces
COM1, COM2 and IEEE-R88 interfaces to
default settings.
3.5.2
Remote Communications
Remote command format to Classic
3.5.2.3
Level (Security)
Reset security level to low
3.5.5.5
Parallel Measurement Mode
Off
3.5.7.3
SDS
All SDS full time ON
3.5.4.2
Summary of Contents for RPM4
Page 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Page 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Page 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Page 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Page 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...