RPM4™ OPERATION AND MAINTENANCE MANUAL
Page
116
The calibration process is performed independently on each Q-RPT to arrive at its optimal fit.
CalTool for RPTs software provided with the RPM4 supports the calibration process of
RPM4 Q-RPTs. CalTool and its documentation are provided on a General Accessories Disk with a
new RPM4 and can be downloaded from www.dhinstruments.com. Most users should use
CalTool software to assist in the calibration of RPM4.
RPM4 is delivered with an interactive Q-RPT calibration utility that steps the operator through
the complete Q-RPT calibration procedure including applying the necessary pressures,
collecting data automatically, calculating new PA and PM values, previewing the results of
the new calibration and activating the results of the new calibration (see the CalTool for RPTs
manual on the General Accessories Disk). RPM4 also provides complete front panel and
remote access to Q-RPT calibration parameters so that Q-RPT calibrations can be performed
without using CalTool software (see Section 5.2.8).
5.2.1.1
PA AND PM COEFFICIENTS
The coefficients used to adjust Q-RPT readings are designated PA (an adder or
offset) and PM (a multiplier or span set). The coefficients affect the Q-RPT
reading following:
Corrected reading = (uncorrected reading • PM) + PA
PA is expressed in units of pressure (always the SI unit, Pascal).
PM is dimensionless.
Each Q-RPT has its own unique PA and PM values. The PA and PM values
currently in use can be viewed and edited from the front panel in the CAL
function (see Section 5.2.7). PA and PM values are automatically edited when
CalTool software results are activated.
As editing PA and PM values will change Q-RPT calibration, they should only
be edited by qualified personnel as part of the calibration process. Caution
should be taken to avoid accidental editing and a security system is
available to prevent access (see Section 3.5.5.5). Incorrect editing of PA
and PM values can cause out of tolerance measurements.
A new RPM4 is delivered with PA and PM values set to zero and 1 for all
ranges. This does not mean that the RPM4 has not been calibrated. In the
original factory calibration, privileged factory coefficients are used for
calibration with the user PA and PM set to zero and 1.
5.2.1.2
AS RECEIVED AND AS LEFT DATA
Frequently, calibration procedures require that as received and as left data be
reported. The necessary information to report as received and as left data on the
calibration of RPM4 Q-RPTs can be obtained in several ways.
When the RPM4 CalTool calibration assistance software is used, as received
data is displayed while running the calibration and is automatically recorded and
reported if desired. As left data is also calculated and presented.
At any time, a) reference pressures applied; b) associated Q-RPT readings;
c) PA and PM and P
offset
values, can be used to calculate as received and as left values.
For example, backing out PA and PM on the as left data yields the Q-RPT
Summary of Contents for RPM4
Page 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Page 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Page 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Page 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Page 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...