To activate the Clean DISPLAY press
[DISPLAY]
,
<7Clean>
. Pressing
<7Clean>
returns to the main run screen with the Clean DISPLAY active.
With the Clean DISPLAY active the main run screen is:
1.
Conventional main run screen first line.
2.
“Clean” second line.
99.135 psi a zHi
To go to a DISPLAY other than Clean, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.7
[HEAD]
PURPOSE
To cause a pressure fluid head correction to be added to or subtracted from the pressure
measured by the RPM4 reference pressure transducer in order to predict the pressure at a
height other than the RPM4’s reference level.
PRINCIPLE
RPM4 measures gauge or absolute pressure at the height of the rear panel
TEST(+)
port.
Frequently, when performing a calibration or test, the device or system under test is at a
different height than the RPM4’s
TEST(+)
port. This difference in height, frequently called
head
, can cause a significant difference between the pressure measured by the RPM4 at its
TEST(+)
port height and the pressure actually applied to the device under test which is at a
different height. In this case, it is useful to make a head correction to the pressure measured
by the RPM4 in order to predict the pressure actually applied at a different height.
RPM4 can accurately determine “head” pressures for gases (nitrogen, helium and air) and
liquids (oil, water) as the pressurized medium. In calculating the head value, standard gravity
(9.80665 m/s
2
) is used. Gas densities are calculated from the selected gas’s standard
density correcting for temperature of 20°C and the measured pressure using the gas’s
compressibility factor to 100 MPa (15 000 psi) and extrapolated above 100 MPa. Oil density
is taken to be 850 kg/m
3
, the density of typical calibration oils at 20°C. Water density is taken
to be 998.2321kg/m
3
(20°C). A custom liquid density may also be specified.
The
[HEAD]
function key is used to specify the height difference between the RPM4
TEST(+)
port and another height. Entering a height of zero turns the function off. The height unit and
the test fluid are specified by pressing
[SPECIAL]
,
<3Head>
(see Section 3.5.3).
Use of the HEAD function to assure in tolerance measurements is most important in low
absolute pressure ranges. Specifying the head height within ± 5 cm (2.5 in.) is adequate
to assure that, even in the worst case, the uncertainty on the head correction will be
insignificant relative to the tolerance on the measurement.
Summary of Contents for RPM4
Page 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Page 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Page 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Page 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Page 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...