RPM4™ OPERATION AND MAINTENANCE MANUAL
Page
58
3.5.1
<1A
UTO
Z>
PURPOSE
To offset the RPM4 Q-RPT(s) readings relative to a reference value in order to compensate
for possible changes in Q-RPT zero between full recalibrations. In differential measurement
mode, to compensate for the difference between the reading of the Hi and Lo Q-RPTs at the
“line pressure” at which differential measurements are made.
To assure operation within measurement uncertainty specifications (see Section 1.2.2), it is
recommended that AutoZ be run (the value of P
offset
updated) whenever vented in gauge mode,
whenever the line pressure is changed in differential mode and at least every 30 days or when
RPM4 has been exposed to temperature changes exceeding 15 ºC (36 ºF) in absolute mode.
PRINCIPLE
AutoZ Purpose and Principle
The main component of the change over time of the RPM4 Q-RPTs is change in zero or
offset, independent of span. Offsetting or “rezeroing” RPM4 Q-RPTs relative to a reference
between recalibrations allows measurement uncertainty specifications to be maintained with
less frequent full calibrations. AutoZ is also used to compensate for the difference between
the Hi and Lo Q-RPT at the “line pressure” at which differential measurements are made.
The RPM4 AutoZero function (AutoZ) provides full on-board support for the rezeroing
process to simplify its application by the user.
The AutoZero function uses three values:
1.
P
std,0
:
The pressure value indicated by the AutoZ reference, the device that is acting as
the reference relative to which to offset the RPM4 Q-RPT.
For Axxx (absolute) Q-RPTs in
absolute measurement
mode, the pressure at which
AutoZ is performed is normally atmospheric pressure and the
P
std,0
value can be supplied
a) by manual entry, b) automatically from a
DHI
RPMx connected to RPM4 COM2
communications port, or c) automatically from an Axxx Lo Q-RPT if one is present in the
RPM4.
For Gxxx or BGxxx (gauge) Q-RPTs or Axxx (absolute) Q-RPTs in
gauge measurement
mode,
P
std,0
is always zero (atmospheric pressure) which is supplied by definition when
the Q-RPT is vented to atmosphere.
For RPM4s with two Q-RPTs of the same type (Axxx, Gxxx, BGxxx) with the Hi Q-RPT in
differential measurement
mode,
P
std,0
is the pressure indicated by the Lo Q-RPT when
the Hi and Lo Q-RPTs are connected to a common pressure.
2.
P
u,0
:
The pressure reading of the Q-RPT, with no AutoZ offset, at the time AutoZ is
performed.
3.
P
offset,G
,
P
offset,A
and P
offset,D
:
The difference between the reading of the Q-RPT with no
AutoZ offset (
P
u,0
) and the indication of the AutoZ reference (
P
std,0
) for gauge (
G
),
absolute(
A
) or differential (
D
) measurement mode:
P
offset
= P
u,0
- P
std,0
P
offset
represents the difference between the Q-RPT reading (
P
u,0
) and the AutoZ standard
(
P
std,0
). The AutoZ function manages the determination, storage and application of
P
offset
to
correct the zero offset in absolute, gauge and differential measurement modes.
When the Q-RPT is calibrated,
P
offset
is set to zero.
P
offset
is then redetermined at regular
intervals using the AutoZ function. The most recent value of
P
offset
is applied to the Q-RPT
reading to correct for change in zero.
Summary of Contents for RPM4
Page 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Page 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Page 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Page 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Page 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...