3. OPERATION
Page
31
The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a
specific test range and setting measurement parameters appropriately for the range. These are
set automatically based on operator entry of desired measurement mode, pressure unit of measure
and range full scale. The selections and settings made by AutoRange are summarized in Table 6.
A range set up using AutoRange and all its range specific settings, may be saved for recall
using
[SETUP]
,
<1range>
(see Section 3.4.1). Saved ranges are available for later selection
using
[RANGE]
(see Section 3.3.1). If a range created by AutoRange is not saved, it is
overwritten by the next AutoRange or erased when another range is selected.
Press [ENT] from the main run screen in RATE mode to use the [AutoTest] function to
AutoRange RPM4 based on DUT tolerance (see Section 3.3.10).
Table 6.
Settings Made by AutoRange
SETTING/SELECTION
PURPOSE
SET TO
SECTION
Unit
Set pressure unit of measure
Operator specified during
AutoRange
3.3.2
Mode
Set pressure measurement mode
(absolute, gauge, negative gauge)
Operator specified during
AutoRange
3.3.3
Full scale
Establish span maximum
Operator specified during
AutoRange
None
AutoRange span
Basis of span based setting
calculations
Absolute mode:
Zero absolute
to AutoRange full scale
Gauge mode:
Zero gauge to
AutoRange full scale
Negative gauge mode
: Minus
atmosphere to AutoRange full
scale (negative equivalent of full
scale on BG15K Q-RPT).
None
Q-RPT
Select most appropriate Q-RPT
to cover range and mode.
Absolute and negative gauge
mode:
Axxx Q-PRT with lowest
default full scale that is greater
than the AutoRange full scale
Gauge mode:
Q-RPT with
lowest span whose full scale is
greater than AutoRange full
scale, Gxxx over Axxx if spans
are equal.
1.2.2.1
Upper limit
Lower limit
Set maximum and minimum
pressure alarm limits.
Upper limit:
105 % of operator
entered full scale pressure
Lower limit:
None in absolute
and gauge modes. Minus
atmosphere in negative gauge
mode, minus equivalent of upper
limit in negative gauge mode on
a BG15K Q-RPT. Minus 66 kPa
(9.6 psi) with BA100K.
3.4.4
Resolution
Set display resolution
10 ppm of AutoRange span or
1 ppm of default Q-RPT span,
whichever is larger.
3.4.2
Stability Limit
Set Ready/Not Ready stability
criterion.
50 ppm of AutoRange span or 2
ppm of default Q-RPT span,
whichever is larger.
3.4.3,
3.3.4
Q-RPT turn down
(Premium class Q-RPTs only)
Turn down Q-RPT
characterization to improve
measurement uncertainty when
span is reduced
See RPM4 pressure
measurement specifications
1.2.2.1
Summary of Contents for RPM4
Page 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VIII N NO OT TE ES S...
Page 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page VI N NO OT TE ES S...
Page 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 114 N NO OT TE ES S...
Page 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 134 N NO OT TE ES S...
Page 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL Page 138 N NO OT TE ES S...