Controller Maintenance
Controller Part Replacement
4
4-37
Access the Electronics Box
Use the following instructions and reference Figure 4-20 to gain access to the electronics
box.
1.
Remove the mains power cord from the power entry module.
2.
Use the instructions in “Chassis and Control Panel” on page 4-9 to remove the
chassis.
3.
Remove the six screws and washers (
) that secure the electronic box cover (
) to
the chassis. The two power supply pca assemblies (items
and
) are attached to
the electronic box cover (
). Disconnect the wiring.
4.
When finished with the part replacement procedure, reconnect the wiring and install
the six screws and washers (
) that secure the electronic box cover (
) to the
chassis.
5.
Use the instructions in “Chassis and Control Panel” on page 4-9 to install the chassis
back into the case.
Power Supply PCA
Use the following instructions and reference Figure 4-20 to replace the power supply pca
(items
and
).
1.
Use the instructions in “Access the Electronics Box” to open the electronics box.
2.
After the cover is removed with the power supply pca assemblies attached, remove
the eight screws and washers (
) that secure the two power supply pca assemblies
(items
and
) to the electronic box cover (
).
3.
Put the two new power supply pca assemblies (items
and
) in the electronic box
cover (
).
4.
Install the eight screws and washers (
) that secure the two power supply pca
assemblies (items
and
) to the electronic box cover (
).
5.
Use the instructions in “Access the Electronics Box” to close the electronics box.
Processor PCA
Use the following instructions and reference Figure 4-20 to replace the processor pca
(
).
1.
Use the instructions in “Access the Electronics Box” to open the electronics box.
2.
Remove the four screws and washers (
) that secure the processor pca (
) to the
main pca standoffs (
).
3.
Install the new processor pca (
) on the main pca standoffs (
). Secure it with four
screws and washers (
).
Summary of Contents for 4322-SYS-NAVAIR
Page 1: ...4322 SYS NAVAIR Automated Pressure Calibrator Service Manual...
Page 10: ...4322 Service Manual vi...
Page 14: ...4322 Service Manual 1 2...
Page 22: ...4322 Service Manual 1 10...
Page 24: ...4322 Service Manual 2 2...
Page 30: ...4322 Service Manual 2 8 hbx064 eps Figure 2 2 Pressure Valve Schematic...
Page 42: ...4322 Service Manual 2 20...
Page 44: ...4322 Service Manual 3 2...
Page 54: ...4322 Service Manual 4 2...
Page 94: ...4322 Service Manual 4 42...
Page 96: ......
Page 106: ...4322 Service Manual 5 12 1 2 3 hbx008 eps Figure 5 2 Intensifier Case Removal...
Page 132: ...4322 Service Manual 6 2...
Page 138: ...4322 Service Manual 7 2...
Page 170: ...4322 Service Manual 7 34...
Page 172: ...4322 Service Manual 8 2...
Page 180: ...4322 Service Manual 9 2...
Page 186: ...4322 Service Manual 9 8...