4322
Service Manual
2-18
Boost pressure selection is dependent on the pressure necessary to complete the UUT
calibration. For a typical UUT calibration, select a set pressure equal to or greater than
the maximum pressure necessary in the calibration sequence. For multiple UUT
calibrations in the same location, set the boost pressure to the lowest available selection
that is equal to or higher than the highest pressure necessary to complete the UUT
calibrations. For example, if the highest pressure in the calibration sequence is 2,500 psi,
select 3,000 psi for the intensifier set pressure. If the highest pressure in the calibration is
6,000 psi, select 7,000 psi for the set pressure.
Note
The higher the set pressure, the faster the Nitrogen Supply depletes.
hbx079.eps
Figure 2-7. Intensifier Schematic (Shown on Front Panel)
Summary of Contents for 4322-SYS-NAVAIR
Page 1: ...4322 SYS NAVAIR Automated Pressure Calibrator Service Manual...
Page 10: ...4322 Service Manual vi...
Page 14: ...4322 Service Manual 1 2...
Page 22: ...4322 Service Manual 1 10...
Page 24: ...4322 Service Manual 2 2...
Page 30: ...4322 Service Manual 2 8 hbx064 eps Figure 2 2 Pressure Valve Schematic...
Page 42: ...4322 Service Manual 2 20...
Page 44: ...4322 Service Manual 3 2...
Page 54: ...4322 Service Manual 4 2...
Page 94: ...4322 Service Manual 4 42...
Page 96: ......
Page 106: ...4322 Service Manual 5 12 1 2 3 hbx008 eps Figure 5 2 Intensifier Case Removal...
Page 132: ...4322 Service Manual 6 2...
Page 138: ...4322 Service Manual 7 2...
Page 170: ...4322 Service Manual 7 34...
Page 172: ...4322 Service Manual 8 2...
Page 180: ...4322 Service Manual 9 2...
Page 186: ...4322 Service Manual 9 8...