Theory of Operation
Controller
2
2-13
Note
The test medium exhausted through the CPS is sent through an additional
coalescing filter. The filter can be identified as an externally visible white
filter which is inserted into the waste bottle when the CPS is mounted into
the Controller. The purpose of this filter is to prevent vaporized test
medium being expelled from the Controller.
Caution
To prevent damage to the Controller, drain as much liquid as
possible out of the UUT before it is connected to the Controller.
Empty the waste container after each test (refer to the
Operators Manual for instructions).