4322
Service Manual
3-4
Required Equipment
The equipment in Table 3-1 is required to complete the calibration procedure.
Table 3-1. Calibration Equipment
Primary Reference
Classifications
Minimum Use Specifications
Suggested Equipment
Automated Pressure
Calibrator (APC)
See Required Components
Supplied with Fluke 4322
Pressure Controller
Range: 0.5 psia to 10,000 psia
(-14.2 to 10,000 psig)
Fluke 4322 required
Supplied with Fluke 4322
CPS module
(Contamination
Prevention System)
Clean and dry.
Supplied with Fluke 4322
Cylinder hose
10,000 psi, Fluke 4322 QC stem
Supplied with Fluke 4322
Intensifier outlet hose
10,000 psi, Fluke 4322 QC stem
Supplied with Fluke 4322
Test hose
10,000 psi, Fluke 4322 QC stem, King
Nutronics QC stem
Supplied with Fluke 4322
Reference pressure
standard
Gauge pressure:
0 to 4 psig: ±0.001 psi
4 to 10,000 psig: ±0.025 % of reading
Vacuum Pressure:
0 to 30 inHg: ±0.025 % of reading or
±0.002 in-Hg, whichever is greater
Absolute Pressure:
0.5 to 35 inHga: ±0.002 inHg
17.2 to 10,000 psia: ±0.025 % of reading
King Nutronics 3689 or
equivalent
Nitrogen Supply
Clean, dry, minimum 500 psig
Nitrogen Cylinder
IEEE cable
Standard: IEEE-488.2
Minimum length: 2 ft
Supplied with Fluke 4322
Environmental Conditions
Laboratory environmental conditions required to complete this procedure:
•
Ambient temperature range: 60 °F to 85 °F. Must not change more than ± 4 °F
throughout the calibration.
•
Ambient relative humidity: 5 % to 95 %. Must not change more than ±20 % RH
throughout the calibration.
•
Low wind and draft area.
Summary of Contents for 4322-SYS-NAVAIR
Page 1: ...4322 SYS NAVAIR Automated Pressure Calibrator Service Manual...
Page 10: ...4322 Service Manual vi...
Page 14: ...4322 Service Manual 1 2...
Page 22: ...4322 Service Manual 1 10...
Page 24: ...4322 Service Manual 2 2...
Page 30: ...4322 Service Manual 2 8 hbx064 eps Figure 2 2 Pressure Valve Schematic...
Page 42: ...4322 Service Manual 2 20...
Page 44: ...4322 Service Manual 3 2...
Page 54: ...4322 Service Manual 4 2...
Page 94: ...4322 Service Manual 4 42...
Page 96: ......
Page 106: ...4322 Service Manual 5 12 1 2 3 hbx008 eps Figure 5 2 Intensifier Case Removal...
Page 132: ...4322 Service Manual 6 2...
Page 138: ...4322 Service Manual 7 2...
Page 170: ...4322 Service Manual 7 34...
Page 172: ...4322 Service Manual 8 2...
Page 180: ...4322 Service Manual 9 2...
Page 186: ...4322 Service Manual 9 8...