OPERATION
ST51A/ST75A/ST75AV Mass Flow Meters
30
Fluid Components International LLC
Changing Flow Units, Example
below lists the steps for making a flow units change with these example parameters: SCFM Flow Units and 3-inch Schedule 40
round pipe size set-up:
Table 7 – Flow Unit Example
Command
Display
Description
ENTER
Menu: >
Starting from Normal Operation Mode.
Z
E for English, M for Metric
After entering Flow Unit Set-up menu (Z), select
English or Metric units.
E
0=SFPS, 1=SCFM, 2=SCFH, 3=LB/H,
4=GPM
After selecting English units (E), select specific
type units. For this example: SCFM (1) which is a
volumetric unit.
1
R Round duct or S rectangular
After selecting Standard Cubic Feet, select round
or rectangular duct.
R
Dia.: 4.0260000
Change? (Y/N)>
After selecting Round duct (R), change displayed
Diameter parameter?
Y
Enter value: #
After responding “Y” to diameter change, enter
diameter value.
3.068
Area: 7.3926572 CMinflow: 0.0000000
Change? (Y/N)>
After entering value for diameter, the computed
area is displayed. Change displayed CMinflow
parameter?
N
Maximum flow: 462.04
Enter to continue
After responding “N” to CMinflow parameter
change, the max. flow is displayed. Press ENTER
to continue.
ENTER
CMaxflow: 462.04
Change? (Y/N)
Change displayed CMaxflow parameter?
Y
Enter value: #
After responding “Y” to CMaxflow change, enter
CMaxflow value. (Note: Instrument checks if
resulting max. flow is too large.)
462.04
CMintemp (F): -40.00000
Change? (Y/N)>
After entering value for CMaxflow, change
CMintemp (F) parameter?
N
CMaxtemp (F): 250.00000
Change? (Y/N)>
After responding “N” to CMintemp change,
change CMaxtemp (F) value?
N
Percent of Range is: OFF
Change to ON?>
After responding “N” to CMaxtemp change,
change Percent of Range On/Off status?
N
100.0 SCFM
After responding “N” to Percent of Range On/Off
status change, instrument returns to Normal
Operation Mode.
V Menu: Configure Outputs (4-20 mA and Source/Sink Outputs)
Use the V menu to configure the analog and source/sink outputs. The flowchart in
below gives an overview of the V menu programming.
Note:
Press [ENTER] as required to loop through all numbered choices. You cannot exit a number loop (Select 1, Select 2,
etc.) without making a valid number choice.
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...