ST51A/ST75A/ST75AV Mass Flow Meters
OPERATION
Fluid Components International LLC
59
ST51A/ST75A/ST75AV Modbus Service Registers
The ST51A/ST75A/ST75AV Modbus supports service registers Totalizer Reset and Totalizer Start/Stop.
●
Reset Command for Totalizer – Use the Function 03 holding command via 4027 integer register to manually reset the instrument’s
totalizer count. This is a write only command.
●
Start/Stop Command for Totalizer – Use the Function 03 holding command via 4026 integer register to manually start or stop the totalizer
count. This is a read/write command.
Table 21 –Modbus Service Data
Service and Setup Functions
Variable/Parameter
Modbus Slave Register
Data Type
Access
Totalizer Reset – To reset the Totalizer
write the value of 0xABCD
4027
Integer
Write
Totalizer Start/Stop
To start the totalizer write 0x1
To stop the totalizer write 0x0
4026
Integer
Read/Write
Totalizer MS
4105
Float
Read
Totalizer LS
4106
Float
Read
Examples of Totalizer Service Register Access using ModScan32
The ST51A Modbus supports enable and reset registers. Using any Modbus master/tester software such as ModScan32 or equivalent, the
user can set parameters.
ModScan32 is a Windows-based utility by WinTECH Software that lets a PC operate as a Modbus master device for testing Modbus
systems. Connect the instrument’s Modbus terminals to one of the host PC’s COM/USB port (a USB connection will require a USB to RS-
485 Serial Adapter).
Note:
Modbus protocol addresses are zero-based, which means the public address values will be offset by “1” relative to
the protocol address value.
Checking the Totalizer Value
1. To read the totalizer value, open
ModScan32 program and set the data
definition (Registers 4105 and 4106) in the
Display Definition dialog box as shown in
figure at right (Pull down Data Definition
from Setup menu or click the Data
Definition icon in the toolbar). (Set Length
value to “2” to include the 2
nd
subsequent
register, 4106.) Click OK when done.
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...