INSTALLATION
ST51A/ST75A/ST75AV Mass Flow Meters
8
Fluid Components International LLC
Figure 5 – Flow Element Installation, Compression Fitting ST51A
The flow element is properly mounted when the tip of the flow element is located 0.50 inches (13 mm) past the pipe centerline. The scale
etched on the side of the insertion pipe indicates the length to the tip of the flow element. Follow the steps below to install the ST51A flow
element.
1. Calculate the insertion depth using the equation below.
I = Insertion depth
I.D. = Pipe inside diameter
T = Pipe wall thickness
C = Pipe mounting coupling and compression fitting (installed length)
𝐼 = 0.50" +
𝐼.𝐷.
2
+ 𝑇 + 𝐶
I = __________
2. Mark the insertion pipe at the calculated insertion depth.
3. Apply proper thread sealant to the tapered pipe thread on the compression fitting and secure into pipe mounting coupling.
4. Insert the flow element to the insertion depth mark making sure the orientation flat is aligned parallel to the flow direction. Hand tighten
the compression nut. Compression fitting manufacturer recommends 1-1/4 turns past hand tight.
5. Tighten the compression nut to the torque specified for the corresponding ferrule material as shown in
Table 2 – Compression Fitting Material
Ferrule
Torque
Teflon
65 in - lbs
316 SST
65 ft - lbs
Note:
The metal ferrule configuration can only be tightened one time. Once tightened, the insertion length is no longer
adjustable.
XXXX
F
L
U
ID
C
OM
P
ON
E
N
T
S
IN
TE
R
N
A
T
IO
N
A
L LLC
y
C00584-1-3
MOUNTING ORIEN TATION OF FLATS
TO BE PARALLEL TO FLOW
2X 1/2" NPT,
OPTIONA L M20 x 1.5
ADJUSTABLE COMPRESSION FITTING,
OPTIONA L TEFLON OR METAL FERRULE
ENCLOSURE MEETS EXPLOSION PROOF
WATER AND DUST TIGHT APPRO VALS
OPTIONA L DUAL LINE DIGI TAL DISPLAY,
90° INCREMENTAL ROTATION
1/2 OR 3/4 INCH NPT
PROCESS CONNECTION
FLOW ARROW
LASER ETCHED INSERTION SCALE
SERIAL NUMBER
REDUCER
BUSHING
PROTECTIVE SHROUD
OVER PROBE TIPS
0.56
[014]
4.00
[102]
U LENGTH
VARIABLE
4.5
[114]
3.28
[83]
3.56
[90]
L
C
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...