Click”C/CE” two times from BAS configuration menu to measurement menu,
you also can continue other operation
Click “C/CE” two times from
measurement mode into configuration
menu
Click“→”to choice sys menu
Lcd Rotate
5.) Scale flow m
3
/h
To change the parameter will lead to the meter output value mutation, if posterior
instrumentation, then modify this parameter should be considered before install posterior
instrumentation( if need it).
Posterior instrumentation-related operational requirement
Steps
Operation instructions
Interface show
Steps
Operation instructions
Interface show
Click”C/CE” to choice confirm and
return Basic configuration menu (click
“
→
” to give up modify.
Meter-scale flow (QMAX) range depending on the caliber meter (DN, unit :mm). Scale flow units:
m
3
/h.
Omin=DN2/3540( the equivalent of the current caliber(0.1m/s velocity)
Qmax = DN2/ 29.5 (equivalent diameter 12m/s velocity)
The scale value of the flow meter relate output and frequency output :
Current output Iout : Instruments measured value / scale flow settings x16 +4
Frequency output Fout: Instruments measured value / scale flow settings values x the frequency
maximum rate settings