Thermographic measurement techniques
3. Measure the radiation intensity (= apparent temperature) from the reflection source
using the following settings:
• Emissivity: 1.0
• D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
Figure 41.3
1 = Reflection source
Figure 41.4
1 = Reflection source
You can not use a thermocouple to measure reflected apparent temperature, because a
thermocouple measures
temperature
, but apparent temperatrure is
radiation intensity
.
41.2.1.1.2
Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
#T810169; r. AN/42241/42268; en-US
119
Summary of Contents for GFx320
Page 1: ...User s manual FLIR GFx3xx series...
Page 2: ......
Page 3: ......
Page 5: ...User s manual FLIR GFx3xx series T810169 r AN 42241 42268 en US v...
Page 6: ......
Page 92: ...Mechanical drawings 25 See next page T810169 r AN 42241 42268 en US 80...
Page 95: ...EU Declaration of conformity 26 See next page T810169 r AN 42241 42268 en US 83...
Page 96: ......
Page 97: ...MET Compliance Data Report truncated 27 See next page T810169 r AN 42241 42268 en US 85...
Page 100: ...IEC IECEE Intertek Test Report truncated 28 See next page T810169 r AN 42241 42268 en US 88...
Page 102: ......
Page 103: ...IEC IECEE Intertek CB Test Certificate 29 See next page T810169 r AN 42241 42268 en US 91...
Page 104: ......
Page 105: ......
Page 106: ...MET Laboratories Test Certificate truncated 30 See next page T810169 r AN 42241 42268 en US 94...
Page 108: ...MET Laboratories Letter of Certification 31 See next page T810169 r AN 42241 42268 en US 96...
Page 163: ......