External optics transmission
, i.e., the optical transmission of any protective windows,
etc. that are set up between the camera and the object of interest.
20.4.3
Recommended values
If you are unsure about the values, the following values are recommended:
Emissivity
0.95
Distance
1.0 m (3.3 ft.)
Reflected appa-
rent temperature
+20°C (+69°F)
Relative humidity
50%
Atmospheric
temperature
+20°C (+69°F)
20.4.4
Procedure
Follow this procedure to change the object parameters globally:
1. Turn the mode wheel to
or
.
2. Push the
button to display a menu.
3. Move the joystick left/right to go to the
Edit
tab.
4. Move the joystick up/down to select
Object parameters
.
5. Push the joystick to display a dialog box.
6. Move the joystick up/down to select the parameter you want to change, then push
the joystick.
7. Move the joystick up/down to change the value, then push the joystick.
8. Push the
button to confirm and leave the setup mode.
Note
• Of the seven parameters above,
emissivity
and
reflected apparent temperature
are
the two most important to set correctly in the camera.
• To change object parameters
locally
, first select a measurement tool in the toolbox,
then select
Use local parameters
. Change the local parameters by selecting
Edit local
parameters
, then edit them in the same way as for global object parameters.
20.4.5
Related topics
• For in-depth information about parameters, and how to correctly set emissivity and re-
flected apparent temperature, see 41
Thermographic measurement techniques
.
#T810169; r. AN/42241/42268; en-US
65
Summary of Contents for GFx320
Page 1: ...User s manual FLIR GFx3xx series...
Page 2: ......
Page 3: ......
Page 5: ...User s manual FLIR GFx3xx series T810169 r AN 42241 42268 en US v...
Page 6: ......
Page 92: ...Mechanical drawings 25 See next page T810169 r AN 42241 42268 en US 80...
Page 95: ...EU Declaration of conformity 26 See next page T810169 r AN 42241 42268 en US 83...
Page 96: ......
Page 97: ...MET Compliance Data Report truncated 27 See next page T810169 r AN 42241 42268 en US 85...
Page 100: ...IEC IECEE Intertek Test Report truncated 28 See next page T810169 r AN 42241 42268 en US 88...
Page 102: ......
Page 103: ...IEC IECEE Intertek CB Test Certificate 29 See next page T810169 r AN 42241 42268 en US 91...
Page 104: ......
Page 105: ......
Page 106: ...MET Laboratories Test Certificate truncated 30 See next page T810169 r AN 42241 42268 en US 94...
Page 108: ...MET Laboratories Letter of Certification 31 See next page T810169 r AN 42241 42268 en US 96...
Page 163: ......