Continuity
Range
OL Reading
Resolution
Accuracy
600.0Ω
660.0Ω
0.1Ω
±(0.9% + 5D)
Continuity: Built-in beeper sounds when measured resist
ance is less than 20Ω and is off when measured
resistance is more than
200Ω. Between 20Ω and 200Ω the beeper will stop at an unspecified point.
Continuity Indicator: 2KHz Tone Buzzer;
Response Time of Buzzer: < 500μsec.
Overload Protection: AC/DC 1000V
Diode
Range
OL Reading
Resolution
Typical Reading
1.500V
1.550V
0.001V
0.400 ~ 0.800V
Open Circuit Voltage: Approx. 1.8V; Overload Protection: AC/DC 1000V
Frequency
Range
OL Reading
Resolution
Accuracy
100.00Hz
100.00Hz
0.01Hz
±(0.1% + 2D)
1000.0Hz
1000.0Hz
0.1Hz
10.000kHz
10.000kHz
0.001kHz
100.00kHz
100.00kHz
0.01kHz
ACV - Minimum Sensitivity (including LoZ ACV):
Range
5Hz ~ 1kHz
1kHz ~ 10kHz
>10kHz
600.0mV
60mV
100mV
Unspecified
6.000V
0.6V
6V
Unspecified
60.00V
6V
10V
Unspecified
600.0V
60V
100V
Unspecified
1000V
600V
Unspecified
Unspecified
ACA - Minimum Sensitivity:
Range
5Hz ~ 10kHz
>10kHz
60.00mA
10mA
Unspecified
600.0mA
60mA
Unspecified
6.000A
2A
Unspecified
10.00A
2A
Unspecified