FLIR DM91 USER MANUAL Document Identifier: DM91-en-US_AB
36
AC Current
Range
OL Reading
Resolution
Accuracy
60.00mA
66.00mA
0.01mA
±(1.5% + 3D)
400.0mA
660.0mA
0.1mA
6.000A
6.600A
0.001A
±(1.5% + 3D)
10.00A
20.00A
0.01A
Accuracy of readings > 10A is unspecified.
Maximum measurement time: > 5A for max. 3 minutes with at least 20-minute rest time.
> 10A for max. 30 seconds with at least 10-minute rest time.
Frequency Response: 45 ~ 1kHz (Sine Wave)
Overload Protection: AC/DC 11A for A terminal. AC/DC 660mA for mA terminal.
DC
μA
Range
OL Reading
Resolution
Accuracy
400.0μA
440.0μA
0.
1μA
±(1.0% + 3D)
4000μA
4400μA
1μA
Input Impedance: approx.
2kΩ
Overload Protection: AC/DC 1000V
AC μA
Range
OL Reading
Resolution
Accuracy
400.0μA
440.0μA
0.
1μA
±(1.0% + 3D)
4000μA
4400μA
1μA
Input Impedance: approx.
2kΩ; Frequency Response: 45 ~ 1kHz (Sine Wave)
Overload Protection: AC/DC 1000V
Resistance
Range
OL Reading
Resolution
Accuracy
600.0Ω
660.0Ω
0.1Ω
±(0.9% + 5D)
6.000kΩ
6.600kΩ
0.001kΩ
±(0.9% + 2D)
60.00kΩ
66.00kΩ
0.01
kΩ
±(0.9% + 2D)
600.0kΩ
660.0kΩ
0.1kΩ
±(0.9% + 2D)
6.000MΩ
6.600MΩ
0.001MΩ
±(0.9% + 2D)
50.00MΩ
55.00MΩ
0.01MΩ
±(3.0% + 5D)
Overload Protection: AC/DC 1000V