FLIR DM91 USER MANUAL Document Identifier: DM91-en-US_AB
38
FLEX Current - Minimum Sensitivity:
Range
5Hz ~ 10kHz
>10kHz
30.00A
3.00A (0.300V)
Unspecified
300.0A
30.0A (0.300V)
Unspecified
3000A
300A (0.300V)
Unspecified
Minimum Frequency: 5Hz
Overload Protection: AC/DC 1000V or 600A
Capacitance
Range
OL Reading
Resolution
Accuracy
1000nF
1100nF
1nF
±(1.9% + 5D)
10.00µF
11.00µF
0.01µF
±(1.9% + 2D)
100.0µF
110.0µF
0.1µF
1.000mF
1.100mF
0.001mF
10.00mF
11.00mF
0.01mF
Overload Protection: AC/DC 1000V
Flex Current
Range
OL Reading
Resolution
Accuracy
30.00A
33.00A
0.01A
±(1.0% + 3D)
300.0A
330.0A
0.1A
3000A
3300A
1A
Accuracy does not include the accuracy of the Flexible Clamp Meter.
Frequency Response: 45 ~ 1kHz (Sine Wave)
Overload Protection: AC/DC 1000V
Type-K Temperature
Range
OL Reading
Resolution
Accuracy
-40.0°C to 400.0°C
440.0°C, -44.0°C
0.1°C
±(1% + 3°C)
-40.0°F to 752.0°F
824.0°F, -44.0°F
0.1°F
±(1% + 5.4°F)
Accuracy does not include the accuracy of the thermocouple probe.
Accuracy specification assumes surrounding temperature stable to ±1 °C. For surrounding temperature
changes of ± 2 °C, rated accuracy applies after 2 hours.
Accuracy specified for use with work light and backlight off only.
Overload Protection: AC/DC 1000V.