Calibration and Verification tests
Operational Verification Procedure (OVP)
Flight 60 Service Manual
|
7-23
7.2.15 OVP Test Results
Serial #:
SW Version:
Hours:
Model:
Tested By:
Date:
Test
Results
Pass
Front Panel self- test
-N/A-
☐
Pressure Sensors Calibration
External pressure gauge …………………………….
Patient Pressure …………………………..
Outlet Pressure …………………………..
☐
Pressure Relief Valve Calibration
Relief Pressure …………………………..
☐
Volume Factor Calibration
…………………………..…………………………………………….
☐
Circuit Test
-N/A-
☐
Volume Verification
Vti ………………………… Vte …………………………
☐
Pressures Verification
External gauge peak Pressure
…………………………..
PPeak …………………………..
☐
PEEP
Set
Tolerance
PEEP Value
Ptrig
☐
5
5 ± 1
10
10 ± 2
20
20 ± 2
Pressure Trigger test (P trig)
-N/A-
☐
Exhalation Valve leak
Pressure level …………………………..
☐
High Pressure Alarm
-N/A-
☐
Power Supply and Battery Charge
-N/A-
☐
Inlet Leak
…………………………………..
☐
Alarm Buzzer Test
-N/A-
☐
O2 Sensor
Set
Tolerance
Value
☐
Air
20.9% - 29%
40%
32% - 48%
60%
52% - 68%
80%
72% - 88%
100%
92% - 100%