Removing and Reinstalling Modules
Replacing the Power Board
Flight 60 Service Manual
|
4-9
4.6.2 Installing the Power Board
To install the power board:
1. Check that the power board sockets do not have any bent pins.
2. Slide the power board through the right and left guide tracks, positioning the
power board connectors over the sockets.
Figure 19 – Power Board Slots and Tracks
Figure 20 – Inserting Power Board into Ventilator Base
3. Firmly push down the power board, making sure that the power board connectors
fit well into the sockets (pin headers) in the base of the ventilator.
Right
Track
Right
Slot
Left
Track
Left
Slot