Calibration and Verification tests
Operational Verification Procedure (OVP)
7-14
| FLIGHT 60 Service Manual
LOW MV
(Low minute ventilation)
Red
High P
(High pressure)
Red
LOW P
(Low pressure)
Red
Audio Paused/Alarm Reset
Orange
On/Off
Green
Panel Lock
Orange
All LEDs between 60 – 100 on the
pressure gauge
light up.
If the test fails:
No LEDs are on -
Check the internal connectors to the front panel.
Some LEDs are on -
Replace the Main Board
7.2.3 Circuit Test
Each time the Exhalation Valve or patient circuit is replaced or serviced, it must
be tested.
To test the patient circuit / exhalation valve:
1. Connect the patient circuit to a 0.5L test lung.
2. Press
Technical
/
, and then press
Circuit TEST
.
3. Press
Enter
/
.
4.
The test is done automatically. It includes the following 3 steps:
Testing the motor
Finding Test Points: 6 calibration points in descending values. The first
value should at least 27.
Verifying: 3 verification points.